spacer.png, 0 kB
Specialised Validation Service
Including generation and execution of IQ, OQ and PQ protocols.


Autoclaves, Sterilizers & Decontamination Equipment
Design, Consultancy, Servicing and Periodic Testing.


Pressure Vessel Design
PD5500
BS3970
European Directive 97/23/EC



Autoclave Control System Refurbishments
Design
Manufacture
Installation
Testing and Validation



Home arrow Logiclave Porous Load Sterilizer
Logiclave Porous Load Sterilizer

Logiclave Porous Load Sterilizers

The Logiclave Porous Load Sterilizer is intended to sterilize porous and non-porous items by a method using saturated and pressurised steam.  It is suitable for sterilisation of air entrained porous items such as linen, dressings, wrapped instruments and utensils, hollow instruments and instruments with lumens.

Nominal sterilization times and temperatures are:
3 minutes at 134 - 137°C
15 minutes at 121 - 124°C

Design

The design of the autoclave is based on the key requirements of the following standards / references:

  • PD5500: 2003
  • BS EN285 1997
  • BS 2646 1993
  • BS 3970 Pt1
  • BS EN 61010
  • HTM 2010
  • C14 NHS Model engineering specifications
  • HSE Guidance Note PM73

In addition, the equipment complies fully with the appropriate requirements of the Pressure Equipment Regulations 1999 (SI 1999 No. 2001) and is CE marked accordingly.

Vessel construction

The chamber, jacket and doors are manufactured from minimum 6mm thick high grade 316L stainless steel for maximum corrosion resistance.  Other process contact materials are generally of conventional copper and brass construction although where appropriate, hygienic stainless steel is also used.  Each stage of the vessel manufacturing process is witnessed and certified by an independent Insurance Company who also verifies the materials used, welder qualifications, weld procedures and acceptance of hydrostatic pressure test.

Control System

The autoclave control system is based on the following hardware components:

  • Rugged industrial PLC type controller
  • Analogue input module
  • Temperature input modules
  • Digital input modules
  • Digital output modules
  • Full colour operator interface (Touch sensitive screen)
  • Independent data recorder (Where specified)

The control system software has been specifically developed for the precision control of the autoclave process.  Program configuration, initiation and monitoring of the process are all possible through the ergonomically designed operator interface.

Data Recording

The Logiclave Porous Load Sterilizer is supplied with a 40 column thermal printer as standard.  Detailed records of each process cycle run including critical times, temperatures and pressures are printed.  Program parameters may also be printed for validation records.  The Logiclave system automatically archives to internal memory all logged cycle data for future retrieval, printing and analysis.  Typically, up to the last twenty cycles are stored in the database.

Further options include independent data recorders with either paper or graphical interfaces provide assurance that sterilizing parameters have been satisfied.

Logiclave Supervisor

A PC based data acquisition system that operates over an industry standard Ethernet connection.  The Logiclave Supervisor package can be connected to any number of Logiclave controllers giving managers the ability to remotely monitor the status of each Logiclave in real time.  All logged data generated by the Logiclave controller is automatically downloaded to a database on the PC.  When reviewing the data, the integrity of the data is verified before being displayed.  

 

Download technical data for Logiclave Porous Load Sterilizer
 
spacer.png, 0 kB